Applications
Biological safety cabinets
are the primary barrier against exposure to aerosols that may be produced
from common microbiology procedures. In addition to protection for personnel
and the room environment, many work programmes require that products are
protected from airborne contamination.BH2000 Series cabinets
provide personnel, environment and product protection in applications
involving biological agents which are classified as being of ordinary,
potential or special hazard. These cabinets are not designed for use with
biological agents of extreme hazard. For applications involving cytotoxic
drugs, CG2000 Series Cytogard cabinets should
be specified.
Australian and International
Standards
Please refer to our Standards
page

Description
BH2000 Series cabinets are
part-recirculating laminar air flow enclosures with high efficiency particulate
air (HEPA) filtration of exhaust air and an air barrier at the work opening.
HEPA-filtered vertical laminar air flow which is recirculated in the work
zone creates an ultra-clean work environment for product protection. An
air barrier between the operator and the work zone is maintained by a flow
of room air into a full width grille in the work opening.
The barrier air mixes with
the laminar flow air in a sump underneath the work surface, and is exhausted
to the room via a HEPA filter. All positive pressure zones and filter seais
are surrounded by negative pressure zones, so as to contain potentially
hazardous aerosols.
Standard cabinets have exhaust
discharge on the right hand side. Optional left hand side or top exhaust
is available. Cabinets are produced with a work zone width of 90cm, 120cm
or 180cm and may be installed on standard laboratory benches or on optional
floor stands. Embodying advanced system-monitoring technology, enhanced
ergonomics and styling, BH2000 Series cabinets set a new standard in safety
cabinet design and performance.
Construction
Cabinet.
Constructed in electro-galvanised
steel with joints welded using a gas-shielded arc process. This method
produces a robust, gas-tight housing which is able to withstand the rigours
of transport and handling. Exterior finish is in a special baked enamel
which has been developed for laboratory equipment.
Work zone.
Constructed in Grade 304
stainless steel with 2B finish. Corners are radiused and crevice-free for
ease of cleaning and all surfaces are carefully dressed to remove sharp
edges. The removable work tray is designed to allow cleaning access to
its underside without removing it from the cabinet. A removable panel provides
cleaning and service access to the rear plenum.A pneumatic mechanism assists
the opening and closing of the viewing window without the need for external
fasteners or catches. The window is self-supporting in the fully-open position
to facilitate cleaning and access for large items. Opening the window with
the cabinet running automatically engages a boost mode for enhanced containment
by activating maximum exhaust airflow and the alarm systems.
Dimensions.
Model |
Exhaust
Discharge |
Overall Dimensions |
Work Zone Dimensions |
Weight |
W |
D |
H |
W |
D |
H |
BHA 90 |
LHS, RHS |
1190 |
765 |
1500 |
870 |
560 |
620 |
210 |
BHA 90 |
Top |
1035 |
765 |
1730 |
870 |
560 |
620 |
210 |
BHA 120 |
LHS, RHS |
1505 |
765 |
1500 |
1180 |
560 |
620 |
245 |
BHA 120 |
Top |
1340 |
765 |
1730 |
1180 |
560 |
620 |
245 |
BHA 180 |
LHS, RHS |
1950 |
765 |
1500 |
1790 |
560 |
620 |
315 |
BHA 180 |
Top |
2105 |
765 |
1730 |
1790 |
560 |
620 |
315 |
All dimensions are in millimeters
(mm)
Note:
1. For
activated-carbon exhaust filter option, width (W) on side-exhaust cabinets
and height (H) on top exhaust cabinets increases by 260mm.
2. Recommended
exhaust discharge clearance is 600mm.
3. Stand
height is nominally 900mm.
Fans.
Separate direct-drive fans
are provided for the exhaust and laminar flow HEPA filters. Fans are speed-controlled
to enable airflows to be maintained through filter life. The fan control
circuits are interlocked so that the laminar flow system will not operate
until the exhaust system has achieved a containment condition.
An audible alarm with rechargeable
battery back-up and visible indicators signify any reduction in airflows.
A manometer gauge located in the work zone monitors fan/filter system pressures.
HEPA filters.
Clyde-Apac Microsealâ„¢ HEPA
filters are manufactured to the most exacting standards for critical applications.
Each filter is individually certified to achieve an arrestance efficiency
of not less than 99.995% to the stringent BS3928 Sodium Flame test. All
testing is conducted in a NATA-registered factory laboratory and a NATA-endorsed
test label, being an extract of the test report, is affixed to each filter.
Electrical and diagnostics.
A low-voltage touch-control
panel is located on the front of the cabinet. A microprocessor-controlled
diagnostic system provides continuous monitoring of critical cabinet functions
with a digital display indicating the nature of any malfunction. The boost
mode and a post-use over-run mode may be manually selected from the control
panel.
An integral fluorescent lamp
housing reduces heat build-up near the operator. Glare-free lamps provide
a minimum lighting intensity of 800 lux at the work floor. A splashproof,
general purpose power outlet is fitted in the work zone.
Quality Assurance and
Control
Please refer to our Standards
page.
Standard features
Low voltage touch
controls
Diagnostics system with digital
display
Alarms and boost mode automatically
engaged when viewing window is open
Boost mode selectable at control
panel
Selectable post-use over-run
timer
Germicidal UV lamp
Pneumatically-assisted viewing
window
Hourmeter to record operating
time
Magnehelic gauge to monitor
fan systems
Splashproof power outlet in
work zone
Stainless steel work opening
cover
Comprehensive operation and
maintenance manual
Options
Additional power outlet
Activated-carbon exhaust filter
Gas tap with solenoid valve
Vacuum tap with filter
Floor stand
Exhaust discharge on left hand
side or top
Fumigation adaptor panels for
work opening and exhaust
Pull-out writing surface (only
with stand)
Experiment timer
On-site testing
BH2000 Series cabinets are
factory-tested and certified by a NATA-registered laboratory. Additional
testing and certification
should be carried out on site prior to use, and subsequently:
After any electrical or mechanical
maintenance
After filter replacement
After re-location
At least annually
In special circumstances, such
as a significant change in the work programme, or where unsafe cabinet
operation is suspected
Laminar
Air Flow Services, a NATA-registered laboratory and a division of Clyde-Apac,
provides comprehensive on-site maintenance, testing and certification services
for safety cabinets, laminar flow work stations, cleanrooms
and HEPA filter installations. This service is available from
fully-equipped, NATA-registered laboratories in major Australian and International
centres.